FOTEMP-OEM
OEM Fiber Optic Signal Conditioner
The FOTEMP-OEM is a small form factor interface module designed for convenient integration of high precision GaAs-based fiber optic temperature sensor technology into OEM applications where common electrical temperature sensors and thermocouples cannot be used.
This OEM interface product is compatible with all TS series fiber optic temperature probes. The sensors are entirely non-metallic and electrically passive, providing immunity to EMI, RFI, NMR, microwaves and radiation is required.
Specifications
Number of channels |
Standard: 1 Additional channels upon request |
Power supply | 9-12 VDC, includes plug-in AC power supply |
Power supply | 350 mA |
Measuring Range | —200°C to +300°C |
Accuracy | +/— 0.2 °C |
Resolution | 0.1 °C |
Sampling rate | 2 Hz |
Analog output |
Options: 0-10 V or 4—20 mA (programmable), BNC |
Digital interface |
Standard: RS-232 Options: RS-485/USB/SPI |
Serial protocol |
ASCII over RS232, etc. See FOTEMP-Assistant data sheet for ASCII protocol details |
Data logging | Supplied with FOTEMP-Assistant software. Excel compatible. |
Optical inteface |
ST, compatible with all TS series fiber optic temperature probes |
Temperature |
Storage: —20°C to +70°C Operating: 0°C to +50°C |
Weight | 80g |
Dimensions | 190 x 100 x 40mm |
Warranty | 2 years |
Calibrations |
Standard -30°C to +300°C Optional -200°C to +300°C |
Calibration Intervals | Under normal use, the instrument does not require recalilbration. When sensors and cables are changed, a One-Point Calibration is required. Recommended cal interval is 12 months for ISO systems. |
Specifications subject to change without notice
Ordering Information
FOTEMP1-OEM
Standard single-channel version with RS232 interface
Other FOTEMP-OEM configurations available upon request. Contact Micronor to discuss your OEM application and requirements.
TS series temperature probes available from stock:
TS2-02 TS2 General Purpose, Length=2m
TS2-06 TS2 General Purpose, Length=6m
TS3-10MM-02 TS3 General Purpose, L1=10mm, L=2m
TS3-10MM-02 TS3 General Purpose, L1=10mm, L=6m
TS4-02 TS4 Harsh Chemicals, Length=2m
TS5-20MM-02 TS5 Medical, L1=20mm, Length=2m
TST2-02 TSTRAFO Transformers, Length=2m
TSNANO-02 TS Nano, Length=2m
Other lengths and configurations available upon request. Contact Micronor to discuss your OEM fiber optic temperature monitoring application and requirements.
Features
- Designed for OEM integration
- Rugged, shielded housing
- Small footprint
- 1-4 channels
- Measuring range: — 200 °C to + 300 °C
- High accuracy: +/— 0.2 °C
- Various serial interface options
- Private label option
- Attractive volume discounts
Applications
- EMI, RFI and microwave environments
- High voltage environments
- Harsh and hazardous environments
- Nuclear environments
- Aerospace applications
- Process monitoring
- Medical applications (MRT)